Best for
- Any time the question is whether an analytical procedure is fit for its intended purpose: designing a validation study, evaluating validation data, verifying a compendial procedure, transferring a procedure to another l…
K-Dense-AI/scientific-agent-skills/skills/analytical-method-validation/SKILL.md
Plan, execute, and document validation, verification, and transfer of analytical procedures under the governing framework - ICH Q2(R2) and Q14, USP <1220>/<1225>/<1226>, ICH M10 bioanalytical, CLSI EP, or ISO/IEC 17025. Use for HPLC, LC-MS/MS, GC, CE, ICP-MS, dissolution, qNMR, qPCR, NIR, and ligand binding or cell-based assays whenever the question is whether a procedure is fit for its intended purpose. Triggers include "method validation", "analytical method validation", "AMV", "validation pro
Decision brief
Plan, execute, and document validation, verification, and transfer of analytical procedures under the governing framework - ICH Q2(R2) and Q14, USP <1220>/<1225>/<1226>, ICH M10 bioanalytical, CLSI EP, or ISO/IEC 17025. Use for HPLC, LC-MS/MS, GC, CE, ICP-MS, dissolution, qNMR, qPCR, NIR, and ligand binding or cell-based assays whenever the question is whet…
Compatibility matrix
| Platform | Status | Evidence | What to check |
|---|---|---|---|
| Codex | Not declared | No explicit evidence | Portability before use |
| Claude Code | Not declared | No explicit evidence | Portability before use |
| Cursor | Not declared | No explicit evidence | Portability before use |
| Gemini CLI | Not declared | No explicit evidence | Portability before use |
Installation
The source command is displayed only when detected. A safe inspection prompt is always available so your agent can explain every action before execution.
npx skills add https://github.com/K-Dense-AI/scientific-agent-skills --skill "skills/analytical-method-validation"Inspect the Agent Skill "analytical-method-validation" from https://github.com/K-Dense-AI/scientific-agent-skills/blob/e7ac42510774624f327003c95b6650e2883bc01d/skills/analytical-method-validation/SKILL.md at commit e7ac42510774624f327003c95b6650e2883bc01d. List every install step, command, network request, credential, file read/write, external action, and rollback step. Explain whether it fits my task. Do not install or execute anything until I approve.
Workflow
Q2(R2) Table 1 decides what is required from the measured attribute, not from the technique. For an assay: specificity, response, accuracy, repeatability, intermediate precision. For a limit test: specificity and DL only. For an identity test: specificity alone. Attributes accep…
Any time the question is whether an analytical procedure is fit for its intended purpose: designing a validation study, evaluating validation data, verifying a compendial procedure, transferring a procedure to another laboratory or instrument, or defending any of these in a repo…
1. Establish which framework governs before designing anything. The same assay validates differently under ICH Q2(R2), USP , ICH M10, CLSI EP, and ISO/IEC 17025. They differ in which characteristics are required, how the studies are laid out, and whether numeric acceptance crite…
This skill plans studies, computes the statistics correctly, and structures the documentation. It does not decide that a procedure is validated, release a batch, accept or reject a run, close an investigation, or substitute for the analyst, the technical reviewer, the quality un…
ICH guidelines are published openly and licensed for reuse with acknowledgement, so their requirements are encoded directly in this skill. USP general chapters, CLSI EP documents, and ISO standards are copyrighted and paywalled. For those, this skill supplies the designation, sc…
Permission review
The documentation asks the agent to run terminal commands or scripts.
python3 plan_validation.py --list-frameworksThe documentation asks the agent to run terminal commands or scripts.
python3 plan_validation.py --framework ich-q2r2 --attribute assay --technique hplc --range-use assayEvidence record
| Signal | Value | Evidence type | Meaning |
|---|---|---|---|
| Quality score | 89/100 | Computed | Documentation, specificity, maintenance, and trust rules |
| Repository stars | 31,966 | Source | Repository attention, not individual Skill quality |
| Compatibility | 0 platforms | Source | Declared in the catalog source record |
| Usage guide | automated source guide | Editorial | Generated or reviewed according to the visible evidence level |
Pinned source
Any time the question is whether an analytical procedure is fit for its intended purpose: designing a validation study, evaluating validation data, verifying a compendial procedure, transferring a procedure to another laboratory or instrument, or defending any of these in a report.
1. Establish which framework governs before designing anything. The same assay validates differently under ICH Q2(R2), USP <1225>, ICH M10, CLSI EP, and ISO/IEC 17025. They differ in which characteristics are required, how the studies are laid out, and whether numeric acceptance criteria are supplied at all. Blending them produces a protocol that satisfies none of them.
2. State acceptance criteria before collecting data. Criteria chosen after seeing results are not acceptance criteria, and deciding them post hoc is a standing audit finding. ICH Q2(R2) deliberately supplies almost no numeric criteria — they have to come from the specification, the analytical target profile (ICH Q14 section 3), or development data. ICH M10 is the exception: it supplies explicit numbers, and they differ between chromatographic assays and ligand binding assays.
This skill plans studies, computes the statistics correctly, and structures the documentation. It does not decide that a procedure is validated, release a batch, accept or reject a run, close an investigation, or substitute for the analyst, the technical reviewer, the quality unit, or the regulator. Every script reports; none of them concludes.
ICH guidelines are published openly and licensed for reuse with acknowledgement, so their requirements are encoded directly in this skill. USP general chapters, CLSI EP documents, and ISO standards are copyrighted and paywalled. For those, this skill supplies the designation, scope, and where to obtain an authorised copy — never the text, never invented thresholds. Do not ask an agent to retrieve, transcribe, or reconstruct their content. If a number matters and it lives in a paywalled document, read it from the authorised copy.
cd skills/analytical-method-validation/scripts
python3 plan_validation.py --list-frameworks
| Key | Governs | Numeric criteria supplied |
|---|---|---|
ich-q2r2 | Release and stability testing of drug substances and products | Almost none — you derive them |
ich-m10 | Bioanalytical concentration measurement (PK, TK, BE) | Yes, and they differ by modality |
usp-1220 | Compendial procedure lifecycle, three stages | Paywalled |
usp-1225 / usp-1226 | Validation / verification of compendial procedures | Paywalled |
clsi | Clinical laboratory measurement procedures (EP series) | Paywalled |
iso-17025 | Lab-developed and modified methods under accreditation | No — "to the extent necessary" |
Q2(R2) replaced Q2(R1) in November 2023 and restructured the characteristics. Range is now the parent characteristic (section 3.2), containing response (linearity) and validation of lower range limits (DL/QL). Accuracy and precision are section 3.3 and may be evaluated in combination against a single criterion. Robustness is treated as a development activity and cross-refers to ICH Q14. Multivariate procedures are addressed explicitly (2.5 and 3.2.2.3), and Annex 2 adds worked examples for techniques Q2(R1) never covered — quantitative ¹H-NMR, NIR, quantitative LC/MS, qPCR, biological assays, and particle size. A Q2(R1)-shaped protocol — a flat list of linearity, range, accuracy, precision, specificity, LOD, LOQ, robustness — is out of date. Note also the error correction dated 30 November 2023 to Table 5 and Tables 6–11.
cd skills/analytical-method-validation/scripts
| Script | Question answered |
|---|---|
plan_validation.py | Which framework, which characteristics, what study layout, what protocol? |
check_response.py | Does the calibration model actually hold across the range? |
check_accuracy_precision.py | What is the recovery, and how much of the variability is between days? |
check_detection_limits.py | What are DL and QL by each allowed approach, and do they serve the reporting threshold? |
check_bioanalytical_run.py | Does this run meet ICH M10 for its modality? |
compare_methods.py | Are two procedures equivalent, at a pre-stated margin? |
All take --format table|tsv|json. Provenance, guideline citations, and caveats go to stderr;
data goes to stdout, so > out.tsv keeps them separate. Exit code is 0 for no findings, 1
when findings were raised, 2 for bad input — so any of them can gate a workflow.
python3 plan_validation.py --framework ich-q2r2 --attribute assay --technique hplc --range-use assay
Q2(R2) Table 1 decides what is required from the measured attribute, not from the technique. For
an assay: specificity, response, accuracy, repeatability, intermediate precision. For a limit
test: specificity and DL only. For an identity test: specificity alone. Attributes accepted include
assay, impurity (quantitative), impurity-limit, and identity.
Reportable range comes from the specification. Q2(R2) Table 2 gives worked examples — 80–120% of declared content for an assay, 70–130% for content uniformity, reporting threshold to 120% of the specification for an impurity.
python3 plan_validation.py --framework ich-q2r2 --attribute impurity --protocol > protocol.md
Every bracketed field is a decision to make and record before data collection. The protocol skeleton deliberately refuses to pre-fill acceptance criteria for Q2(R2) work, because there is no defensible default.
python3 check_response.py -i calibration.csv --max-back-calc-error 2
Input is level,response, one row per injection; repeated rows at the same level are replicates,
and supplying them is what makes the linearity test possible.
Real output from a curve that a coefficient of determination would wave through:
statistic value
distinct levels 5
slope 166.6000
intercept 2495.0000
intercept CI includes 0 no
coefficient of determination (r2) 0.9830
lack-of-fit F 469.5294
lack-of-fit p 1.5139e-06
runs test p 0.0492
level n mean_response mean_back_calculated relative_error_pct
50.0000 2 10075.0000 45.4982 -9.0036
75.0000 2 15150.0000 75.9604 1.2805
100.0000 2 20050.0000 105.3721 5.3721
125.0000 2 24050.0000 129.3818 3.5054
150.0000 2 26450.0000 143.7875 -4.1417
r² = 0.983 and the model is unusable: −9.0% back-calculated error at the bottom of the range, lack-of-fit p = 1.5 × 10⁻⁶, non-random residual signs. r² is not evidence of linearity — it rises with range and is nearly insensitive to curvature. The lack-of-fit F test against pure error and the residual pattern are the evidence, which is why Q2(R2) 3.2.2.1 asks for an analysis of the deviation of points from the line rather than a correlation coefficient alone.
Add --weight 1/x2 for a wide-range curve. The script flags heteroscedasticity when the residual
variance in the top third of the range exceeds the bottom third by more than 10×, because an
unweighted fit then biases exactly the low end where a reporting threshold lives.
python3 check_accuracy_precision.py -i ap.csv --accuracy-limit 2 --rsd-limit 1.0 --design-check assay
Input is level,measured,group, where group is the intermediate-precision factor — day, analyst,
or instrument.
level component sd rsd_pct df ci90_low_sd ci90_high_sd
100 repeatability (within group) 0.0707 0.0707 3 0.0438 0.2065
100 between-group 1.6515 1.6515 2 n/a n/a
100 intermediate precision (total) 1.6530 1.6530 2.0037 0.9554 7.2821
Repeatability of 0.07% RSD looks superb; intermediate precision is 1.65%, twenty-three times larger, because the variability lives entirely between days. Reporting the within-day figure as the procedure's precision would understate routine performance by more than an order of magnitude. This is why the script fits a one-way random-effects model rather than pooling.
Two traps the script handles for you:
--require-ci-within-limit enforces that the whole confidence interval sits inside the
limit, not just the mean. Q2(R2) 3.3.1.4 asks for the interval to be compatible with the
criterion; a mean that scrapes inside on six replicates has not demonstrated much.python3 check_detection_limits.py --calibration lowcal.csv --blanks blanks.csv \
--confirm-ql 0.05 --confirm-data ql_check.csv --reporting-threshold 0.05
approach sigma slope DL QL
sd-and-slope (sigma = residual SD of regression) 7.2816 5033.3490 0.0048 0.0145
sd-and-slope (sigma = SD of y-intercept) 4.3303 5033.3490 0.0028 0.0086
sd-and-slope (sigma = SD of 8 blanks) 3.7702 5033.3490 0.0025 0.0075
The same data give QL estimates spanning 1.9×, purely from the choice of σ. Q2(R2) 3.2.3.5
therefore requires the limit and the approach used to determine it to be reported, and an
estimated limit to be confirmed with samples at or near it. For an impurity procedure the QL must
be at or below the reporting threshold. Reaching for 3.3σ/slope reflexively, reporting one number
with no named approach, and never confirming it are three separate findings.
python3 check_bioanalytical_run.py --modality chromatographic --run run1.csv
python3 check_bioanalytical_run.py --modality lba --isr isr.csv
python3 check_bioanalytical_run.py --modality lba --criteria
--modality is mandatory and has no default, because the criteria genuinely differ:
| Chromatographic | Ligand binding assay | |
|---|---|---|
| Calibration tolerance | ±15%, ±20% at LLOQ | ±20%, ±25% at LLOQ and ULOQ |
| Accuracy / precision | ±15% / ≤15% CV (±20% / ≤20% at LLOQ) | ±20% / ≤20% CV (±25% / ≤25% at LLOQ and ULOQ) |
| A&P design | 4 QC levels, 5 replicates/run, ≥3 runs over ≥2 days | 5 QC levels, 3 replicates/run, ≥6 runs over ≥2 days |
| Total error | no such criterion | ≤30%, ≤40% at LLOQ and ULOQ |
| ISR agreement | ±20% for ≥2/3 of repeats | ±30% for ≥2/3 of repeats |
Applying the ±15% chromatographic numbers to a ligand binding assay, or importing the LBA total-error criterion into a chromatographic method, are both common and both wrong.
The run check enforces the per-level rule that gets missed: at least 2/3 of all QCs and at least 50% at each level. A run can pass the overall fraction while a single level fails completely.
finding: QC level high: 0/2 within tolerance (0%); M10 requires at least 50% at each level
python3 compare_methods.py -i paired.csv --margin 2 --relative --slope-tolerance 0.05
mean difference (%) 1.4646
TOST margin 2.0000
TOST p-value 1.0528e-13
90% CI (TOST) 1.44127 to 1.48797
equivalent at stated margin yes
--- for contrast only ---
paired t-test p (NOT equivalence) 0.0000
OLS slope (biased here) 1.0396
Deming slope 1.0398
Passing-Bablok slope 1.0351
Two errors this replaces:
The script also flags proportional bias — when the difference trends with concentration, a single mean bias and its limits of agreement are misleading regardless of how tight they look.
references/framework-selection.md — which framework governs, and the questions that decide itreferences/ich-q2r2.md — structure, Table 1 and Table 2, per-characteristic recommended datareferences/ich-m10-bioanalytical.md — the full chromatographic and LBA criteria side by sidereferences/compendial-and-clsi.md — USP, CLSI and ISO designations, scope, and how to cite themreferences/statistics.md — the statistical methods, why each one, and the common errorsreferences/source-ledger.md — provenance and research dates for every claim in this skillassets/validation-protocol-template.md — protocol structure with criteria stated up frontassets/validation-report-template.md — report structure with raw-data traceabilityAlternatives
coreyhaines31/marketingskills
When the user wants to plan, design, or implement an A/B test or experiment, or build a growth experimentation program. Also use when the user mentions "A/B test," "split test," "experiment," "test this change," "variant copy," "multivariate test," "hypothesis," "should I test this," "which version is better," "test two versions," "statistical significance," "how long should I run this test," "growth experiments," "experiment velocity," "experiment backlog," "ICE score," "experimentation program
event4u-app/agent-config
Use BEFORE writing/changing tests, adding mocks, or test-only methods on production classes — vs mocking-the-mock, production pollution, partial mocks, and overfit/tautological assertions
event4u-app/agent-config
Use when the user says "review the design", "check the UI", or wants a comprehensive UI/UX review. Uses a 7-phase methodology covering interaction, responsiveness, accessibility, and more.
event4u-app/agent-config
Use when writing, generating, or improving Pest tests for Laravel — clear intent, good coverage, maintainable structure, and alignment with project testing conventions.